Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 17: (1 year) CAL VER /5500 INSTRUMENT: Fluke 19: (1 year) CAL VER /5500 DATE: 22-Jun-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 64 NUMBER OF LINES: 262 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Preliminary Test Plan, 4/6/98 # # Compatibility: # 5500/CAL or MET/CAL version 5.1 and later # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Subprocedures: # None # # Fluke make no warranty, expressed or implied, as to the fitness or # useability of this procedure. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP No warm-up time requirement. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 90%. 1.006 HEAD {DISPLAY TEST} 1.007 MESS Hold down the RANGE button while turning the UUT rotary 1.007 MESS function switch from OFF to VDC. 1.007 MESS 1.008 EVAL -s "All Segments" : Are all segments illuminated? 2.001 MESS 2.002 ASK- U 2.003 HEAD {ROTARY SWITCH TESTS} 2.004 DISP Release the RANGE button and turn the switch to {VAC}. 2.005 MATH MEM1 = 0 2.006 MEMI Enter UUT reading: 2.007 MEME 2.008 MEMC display 12U 3.001 DISP Rotate the UUT function switch to {VDC}. 3.002 MATH MEM1 = -32 3.003 MEMI Enter UUT reading: 3.004 MEME 3.005 MEMC display 12U 4.001 DISP Rotate the UUT function switch to {mVDC}. 4.002 MATH MEM1 = -64 4.003 MEMI Enter UUT reading: 4.004 MEME 4.005 MEMC display 12U 5.001 DISP Rotate the UUT function switch to {Ohms}. 5.002 MATH MEM1 = -96 5.003 MEMI Enter UUT reading: 5.004 MEME 5.005 MEMC display 12U 6.001 JMPT 6.004 Fluke 19: (1 year) CAL VER /5500 6.002 DISP Rotate the UUT function switch to {mA}. 6.003 JMP 6.005 6.004 DISP Rotate the UUT function switch to {A}. 6.005 MATH MEM1 = -160 6.006 MEMI Enter UUT reading: 6.007 MEME 6.008 MEMC display 12U 7.001 DISP Rotate the UUT function switch to {Diode}. 7.002 MATH MEM1 = -128 7.003 MEMI Enter UUT reading: 7.004 MEME 7.005 MEMC display 12U 8.001 ASK+ U 8.002 HEAD {AC VOLTAGE TESTS} 8.003 DISP Rotate the UUT function switch to OFF then VAC. 8.004 DISP Connect the 5500A to the UUT as follows: 8.004 DISP [32] 5500A NORMAL HI to UUT Volt-Ohm-Diode Input 8.004 DISP [32] 5500A NORMAL LO to UUT COM Input 8.005 HEAD AC VOLTAGE TESTS: {400mV Range} 8.006 5500 0.0V S 2W 8.007 MEMI Enter UUT reading in volts: 8.008 MEME 8.009 MEMC V 0.9U 9.001 5500 400 30.0mV 0.5U 45H SI 2W 10.001 5500 400 350.0mV 3.4U 45H SI 2W 11.001 5500 400 350.0mV 3.4U 1kH SI 2W 12.001 5500 400 350.0mV 8.6U 10kH SI 2W 13.001 5500 400 350.0mV 8.6U 20kH SI 2W 14.001 HEAD AC VOLTAGE TESTS: {4V Range} 14.002 5500 4 3.500V 0.034U 45H SI 2W 15.001 5500 4 3.500V 0.034U 1kH SI 2W 16.001 5500 4 3.500V 0.086U 10kH SI 2W 17.001 5500 4 3.500V 0.086U 20kH SI 2W 18.001 HEAD AC VOLTAGE TESTS: {40V Range} 18.002 5500 40 35.00V 0.34U 45H SI 2W 19.001 5500 40 35.00V 0.34U 1kH SI 2W 20.001 5500 40 35.00V 0.86U 10kH SI 2W 21.001 5500 40 35.00V 0.86U 20kH SI 2W 22.001 HEAD AC VOLTAGE TESTS: {400V Range} 22.002 5500 400 350.0V 3.4U 45H SI 2W 23.001 5500 400 350.0V 3.4U 1kH SI 2W 24.001 HEAD AC VOLTAGE TESTS: {1000V Range} 24.002 5500 1000 1000V 17U 45H SI 2W 25.001 5500 1000 1000V 17U 1kH SI 2W 26.001 HEAD {FREQUENCY TESTS} 26.002 DISP Press the RANGE button to lock the 400mV Range. 26.002 DISP Press the Hz button to choose the Frequency mode. 26.003 HEAD FREQUENCY TESTS: {19.999kHz Range} 26.004 5500 20 19.000kH 0.003U 100mV SI 2W 27.001 HEAD FREQUENCY TESTS: {199.99kHz Range} 27.002 5500 200 190.00kH 0.03U 130mV SI 2W 28.001 HEAD FREQUENCY TESTS: {199.99Hz Range} 28.002 5500 200 3.00H 0.01U 120mV SI 2W 29.001 HEAD {DUTY CYCLE: 4V Range} 29.002 DISP Press the % button to choose the Duty Cycle mode. 29.002 DISP Press the RANGE button to lock the 4V Range. 29.003 M550 2.0V 1.7Voff 29.004 5500 50pct 17U 100H TI 2W #! T.U.R. not calculated because tolerance or floor unspecified. 30.001 HEAD {DUTY CYCLE: 40V Range} 30.002 DISP Press the RANGE button to lock the 40V Range. 30.003 M550 8.0V 4.0Voff 30.004 5500 50pct 17U 100H TI 2W #! T.U.R. not calculated because tolerance or floor unspecified. 31.001 M550 * 31.002 HEAD {DC VOLTAGE TESTS} 31.003 DISP Rotate the UUT function switch to VDC. 31.004 HEAD DC VOLTAGE TESTS: {4V Range} 31.005 5500 4 3.500V 0.008U 2W 32.001 HEAD DC VOLTAGE TESTS: {40V Range} 32.002 5500 40 -35.00V 0.08U 2W 33.001 HEAD DC VOLTAGE TESTS: {400V Range} 33.002 5500 400 350.0V 0.8U 2W 34.001 HEAD DC VOLTAGE TESTS: {1000V Range} 34.002 5500 1000 1000V 3U 2W 35.001 HEAD {DC MILLIVOLT TEST} 35.002 DISP Rotate the UUT function switch to mVDC. 35.003 HEAD DC MILLIVOLT TEST: {400mV Range} 35.004 5500 400 0.0mV 0.1U 2W 36.001 5500 400 350.0mV 0.8U 2W 37.001 HEAD {DIODE TESTS} 37.002 DISP Rotate the UUT function switch to Diode. 37.003 5500 1kZ S 2W 37.004 ACC 0.75V TOL 37.005 MEMI Enter UUT reading in volts: 37.006 MEME 37.007 MEMC V 0.3U 38.001 5500 3.000V 0.055U 2W 39.001 HEAD {RESISTANCE TESTS} 39.002 DISP Rotate the UUT function switch to Ohms. 39.003 DISP Connect the 5500A to the UUT as follows: 39.003 DISP [32] 5500A AUX HI to UUT Volt-Ohm-Diode Input 39.003 DISP [32] 5500A AUX LO to UUT COM Input 39.003 DISP [32] 5500A NORMAL HI to UUT Volt-Ohm-Diode Input 39.003 DISP [32] 5500A NORMAL LO to UUT COM Input 39.003 DISP Connect the sense leads to the UUT first. 39.004 HEAD RESISTANCE TESTS: {400 Ohm Range} 39.005 5500 400 0.0Z 0.3U CW 40.001 5500 400 350.0Z 1.2U CW 41.001 HEAD RESISTANCE TESTS: {4 kOhm Range} 41.002 5500 4 3.500kZ 0.020U CW 42.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 42.002 5500 40 35.00kZ 0.20U CW 43.001 HEAD RESISTANCE TESTS: {400 kOhm Range} 43.002 DISP Connect the 5500A to the UUT as follows: 43.002 DISP [32] 5500A NORMAL HI to UUT Volt-Ohm-Diode Input 43.002 DISP [32] 5500A NORMAL LO to UUT COM Input 43.003 5500 400 350.0kZ 2.0U 2W 44.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 44.002 5500 4 3.500MZ 0.020U 2W 45.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 45.002 5500 40 35.00MZ 0.43U 2W #! WARNING: Test Tol 4.3e+005, Sys Tol 1.455e+005, TUR 2.9553 (< 4.00). 46.001 HEAD {CONDUCTANCE TEST} 46.002 DISP Press the Range button twice for 40nS range. 46.003 HEAD CONDUCTANCE TEST: {40nS Range} 46.004 5500 40 10.00nY 0.20U 2W 47.001 EVAL -s "OPEN" : Does UUT indicate between -0.01 and 0.05? 48.001 HEAD {CAPACITANCE TESTS} 48.002 DISP Press the BLUE button to select Capacitance. 48.003 HEAD CAPACITANCE TESTS: {5.00uF Range} 48.004 5500 5.00 4.70uF 0.06U 2W #! WARNING: Test Tol 6e-008, Sys Tol 2.222e-008, TUR 2.7003 (< 4.00). 49.001 HEAD CAPACITANCE TESTS: {0.500uF Range} 49.002 5500 0.5 0.470uF 0.006U 2W #! WARNING: Test Tol 6e-009, Sys Tol 1.893e-009, TUR 3.1696 (< 4.00). 50.001 HEAD CAPACITANCE TESTS: {0.0500uF Range} 50.002 TOL -0.0006U +0.0011U 50.003 5500 0.05 0.0470uF TOL 2W #! WARNING: Test Tol 6e-010, Sys Tol 1.893e-010, TUR 3.1696 (< 4.00). 51.001 HEAD CAPACITANCE TESTS: {5.00nF Range} 51.002 5500 5.00 4.70nF -0.06U +0.51U 2W #! WARNING: Test Tol 6e-011, Sys Tol 2.786e-011, TUR 2.1536 (< 4.00). 52.001 DISP Disconnect the 5500A from the UUT. 52.002 EVAL -s "OPEN" : Does UUT indicate between 0.30 and 0.45? 53.001 JMPT 55.002 Fluke 19: (1 year) CAL VER /5500 # Fluke 17 53.002 TARGET 53.003 HEAD {DC MILLIAMP TESTS} 53.004 DISP Rotate the UUT function switch to mA. 53.005 DISP Connect the 5500A to the UUT as follows: 53.005 DISP [32] 5500A AUX HI to UUT mA Input 53.005 DISP [32] 5500A AUX LO to UUT COM Input 53.006 HEAD DC MILLIAMP TESTS: {40mA Range} 53.007 5500 40 35.00mA 0.38U 2W 54.001 HEAD DC MILLIAMP TESTS: {400mA Range} 54.002 5500 400 350.0mA 3.8U 2W 55.001 JMP 57.001 # Fluke 19 55.002 TARGET 55.003 HEAD DC AMP TESTS: {4000mA Range} 55.004 DISP Rotate the UUT function switch to A. 56.001 DISP Connect the 5500A to the UUT as follows: 56.001 DISP [32] 5500A AUX HI to UUT A Input 56.001 DISP [32] 5500A AUX LO to UUT COM Input 55.005 5500 4000 3500mA 38U 2W 56.002 HEAD DC AMP TESTS: {10A Range} 56.003 5500 10 10.00A 0.12U 2W 57.001 DISP Press the BLUE button to select AC. 57.002 JMPT 61.002 Fluke 19: (1 year) CAL VER /5500 # Fluke 17 57.003 TARGET 57.004 HEAD {AC MILLIAMP TESTS} 57.005 HEAD AC MILLIAMP TESTS: {40mA Range} 57.006 5500 40 35.00mA 0.48U 45H SI 2W 58.001 5500 40 35.00mA 0.48U 2kH SI 2W 59.001 HEAD AC MILLIAMP TESTS: {400mA Range} 59.002 5500 400 350.0mA 4.8U 45H SI 2W 60.001 5500 400 350.0mA 4.8U 2kH SI 2W #! WARNING: Test Tol 0.0048, Sys Tol 0.00275, TUR 1.7455 (< 4.00). 61.001 JMP 64.001 # Fluke 19 61.002 TARGET 61.003 HEAD {AC AMP TESTS} 61.004 HEAD AC AMP TESTS: {4000mA Range} 61.005 5500 4000 3500mA 48U 45H SI 2W 62.001 HEAD AC AMP TESTS: {10A Range} 62.002 5500 10 10.00A 0.15U 45H SI 2W 63.001 5500 10 10.00A 0.15U 1kH SI 2W 64.001 END #! T.U.R.s less than 4.00: 6 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 2 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.